Chapter 4 PAS Application Software
The available options are shown in the following table.
In pulse mode, either leading, or
trailing edge of the input pulse is
recognized as an event. In KYZ
mode, both leading and trailing
edges of the input pulse are
recognized as separate events.
NORMAL (N.O.),
INVERTING (N.C.)
For the normal polarity, the open
to closed transition is considered a
pulse. For the inverting polarity,
the closed to open transition is
considered a pulse.
It has no meaning in KYZ mode
where both transitions are used.
The amount of time while the state
of the digital input should not
change to be recognized as a new
state. Too low debounce time could
produce multiple events on the
input change.
The debounce time is used the same for both digital inputs. If
you change the debounce time for one digital input, the same
debounce time is automatically assigned to the other.
Using Relay Outputs
The PM175 is equipped with two relays. Each relay can be
operated either locally from the alarm/control setpoints in
response to an external event, or by a remote command sent
through communications, and can also be linked to an internal
pulse source to produce energy pulses.
Comments to this Manuals